Component: nanoElute system (SN1837934487616) Framework: nanoElute V1.1.0.63 (356, 2109) Simulation: False --------------------------------------------------------------- Component: Gradient pump Driver: 0.9.9.25 Firmware: 182 Hardware: 7 Serial number: 219027 Flow offset A: -0.0015 µL/min Flow factor A: 1.026452 Flow offset B: 0.0027 µL/min Flow factor B: 2.3 Pressure offset A: 125.5967 Pressure factor A: 0.01972 Pressure offset B: 130.9842 Pressure factor B: 0.01972 Valve shifts A: 34346 Rotor shifts A: 3519 Syringe mileage A: 332777 mm Seal mileage A: 152010 mm Valve shifts B: 32794 Rotor shifts B: 19792 Syringe mileage B: 356188 mm Seal mileage B: 167263 mm --------------------------------------------------------------- Component: Autosampler Driver: 2.3.2 (2.3.16075.643) Firmware: 2.2.9 (2.2.17045.1553) --------------------------------------------------------------- Component: Journal --- Preparation: 04/25/23 12:46:35 Preparation A [µL]: 7.76769255575588 Preparation B [µL]: 16.6607036000276 --- Leak test: 07/06/23 14:06:05 LT: Pump A head [µL/min]: 0.113516214549691 LT: Pump B head [µL/min]: 1.51676093315325 LT: Route 1 tubing [µL/min]: -0.000234042894840241 LT: Route 2 tubing [µL/min]: 0.0035433333158493 LT: Route 3 with loop [µL/min]: 0.000482999989986402 --- Flow restriction test: 07/06/23 14:23:04 FR: Route 1 [µL/min]: 0.925296932170391 FR: Route 2 [µL/min]: 2.72746394351006 FR: Route 3 with loop [µL/min]: 4.04720909586467 ---------------------------------------------------------------