Component: nanoElute system (SN1837934487616) Framework: nanoElute V1.1.0.63 (356, 2109) Simulation: False --------------------------------------------------------------- Component: Gradient pump Driver: 0.9.9.25 Firmware: 182 Hardware: 7 Serial number: 219027 Flow offset A: -0.0015 µL/min Flow factor A: 1.026452 Flow offset B: 0.0021 µL/min Flow factor B: 2.3 Pressure offset A: 131.0157 Pressure factor A: 0.01972 Pressure offset B: 127.9315 Pressure factor B: 0.01972 Valve shifts A: 26946 Rotor shifts A: 1322 Syringe mileage A: 275753 mm Seal mileage A: 94986 mm Valve shifts B: 25533 Rotor shifts B: 12531 Syringe mileage B: 292408 mm Seal mileage B: 103483 mm --------------------------------------------------------------- Component: Autosampler Driver: 2.3.2 (2.3.16075.643) Firmware: 2.2.9 (2.2.17045.1553) --------------------------------------------------------------- Component: Journal --- Leak test: 06/22/22 11:00:55 LT: Pump A head [µL/min]: 0.190171075297258 LT: Pump B head [µL/min]: 0.318348055563774 LT: Route 1 tubing [µL/min]: -4.2438576221466E-05 LT: Route 2 tubing [µL/min]: 0.0106099998235703 LT: Route 3 with loop [µL/min]: -0.00514433322668076 --- Flow restriction test: 06/28/22 14:12:57 FR: Route 1 [µL/min]: 0.886887109298706 FR: Route 2 [µL/min]: 2.69389727752686 FR: Route 3 with loop [µL/min]: 4.3873388604934 --- Preparation: 07/11/22 07:51:58 Preparation A [µL]: 7.6026542216873 Preparation B [µL]: 16.5151431404113 ---------------------------------------------------------------